Application of Topology Optimization in Design of Stem Profile in Hip Implants Using Finite Element Method
Subject Areas : optimization and simulationMohammad Reza Niroomand 1 , Farzad Boroomand 2
1 - Department of Mechanical Engineering, Payame Noor University, Iran
2 - Department of Mechanical Engineering, Najafabad Branch, Islamic Azad University, Isfahan, Iran
Keywords: Stem Profile Design, Topology Optimization, Hip Implant, Finite Element, Stress Analysis,
Abstract :
Appropriate design of stem shape is an important factor in total hip replacement. It affects the attachment of the implant to the bone and the stability of the implant. Using topology optimization, this study has been an attempt to propose an optimized model of the stem profile of the hip implants. In this regard, a three-dimensional finite element model of the implant has been combined with a stiffness-based topology optimization algorithm to reduce the relative motion between the implant and the bone. The objective function in the optimization problem is the compliance of the structure which should be minimized. Also, a constraint on usable volume is applied to the structure. Then smoothing process has been done on the optimal model to prevent its geometric complexities. Results show that the final model has a smaller weight, less displacement, and more uniform stress distribution. In addition, using conventional production methods, this model can be easily produced.
[1] Kayabasi, O., Erzincanli, F., Finite Element Modelling and Analysis of a New Cemented Hip Prosthesis, Advances in Engineering Software, Vol. 37, No. 7, 2006, pp. 477-483.
[2] Bennett, D., Goswami, T., Finite Element Analysis of Hip Stem Designs, Materials & design, Vol. 29, No. 1, 2008, pp. 45-60.
[3] Sabatini, A., Goswami, T., Hip Implants VII: Finite Element Analysis and Optimization of Cross-Sections, Materials & Design, Vol. 29, No. 7, 2008, pp. 1438-1446.
[4] Wang, M., Optimization of Femoral Prosthesis Based on Comprehensive Evaluation of Structure and Material Properties, Journal of Mechanics in Medicine and Biology, Vol. 16, No. 8, 2016, pp. 1640013.
[5] Boyle, C., Kim, I. Y., Comparison of Different Hip Prosthesis Shapes Considering Micro-Level Bone Remodeling and Stress-Shielding Criteria Using Three-Dimensional Design Space Topology Optimization, Journal of Biomechanics, Vol. 44, No. 9, 2011, pp. 1722–1728.
[6] Fraldi, M., Esposito, L., Perrella, G., Cutolo, A. and Cowin, S., Topological Optimization in Hip Prosthesis Design, Biomechanics and Modeling in Mechanobiology, Vol. 9, No. 4, 2010, pp. 389–402.
[7] Virulsri, C., Tangpornprasert, P. and Romtrairat, P., Femoral Hip Prosthesis Design for Thais Using Multi-Objective Shape Optimization, Materials and Design, Vol. 68, 2015, pp. 1-7.
[8] Katoozian, H., Davy, D. T., Three-Dimensional Shape Optimization of Femoral Component of Hip Prostheses with Nonlinear Frictional Interface, Journal of Mechanics in Medicine and Biology, Vol. 4, No. 2, 2004, pp. 161-172.
[9] Tanino, H., Higa, H. I. M., Omizu, N., Nishimura, I., Matsuda, K., Mitamura, Y., and Matsuno, T., Three-Dimensional Computer-Aided Design Based Design Sensitivity Analysis and Shape Optimization of the Stem Using Adaptive P-Method, Journal of Biomechanics, Vol. 39, No. 10, 2006, pp. 1948-1953.
[10] Sazesh, S., Ghassemi, A, Ebrahimi, R., and Khodaei, M., Experimental and Numerical Analysis of Titanium/HA FGM for Dental Implantation, International Journal of Advanced Design and Manufacturing Technology, Vol. 10, No. 1, 2017, pp. 57-74.
[11] Rahmati, S., Kheirollahi, H., and Azari, A., Design and Analysis of a New Dental Implant using Finite Element Method, International Journal of Advanced Design and Manufacturing Technology, Vol. 6, No. 3, 2013, pp. 23-32.
[12] Munckhof, S. V. D., Sadri Nikooyan, A., and Zadpoor, A. A., Assessment of Osteoporotic Femoral Fracture Risk: Finite Element Method as a Potential Replacement for Current Clinical Techniques, Journal of Mechanics in Medicine and Biology, Vol. 15, No. 3, 2015, pp. 1530003.
[13] Jagtap, K. A., Pawade, R. S., and Giradkar, K. V., Investigations on Surface Integrity and Electrochemical Behavior of Machined Co-Cr-Mo Bio-implant Alloy, International Journal of Advanced Design and Manufacturing Technology, Vol. 9, No. 4, 2016, pp. 51-58.
[14] Xiaojian, W., Shanqing, X., Shiwei, Z., Wei, X., Martin, L., Peter, C., Qian, M., Milan, B., and Yi, M. X., Topological Design and Additive Manufacturing of Porous Metals for Bone Scaffolds and Orthopedic Implants: A Review, Biomaterials, Vol. 83, 2016, pp. 127-141.
[15] Shiwei, Z., Wei, L., Guangyong, and Qing, L., A Level-Set Procedure for the Design of Electromagnetic Metamaterials, Optics Express, Vol. 18, No. 7, 2010, pp. 6693-6702.
[16] Chih Ling, C., Chen Sheng, C., Chang Hung, H., and Hsu, H., Finite Element Analysis of the Dental Implant Using a Topology Optimization Method, Medical Engineering and Physics, Vol. 34, No. 7, 2012, pp. 999-1008.